AP-200900-A

INJECTOMAT AGILIA, THE NEW GENERATION OF STAND ALONE SYRINGE PUMPS 

• flow rate range 0.1 - 1200 ml/h in normal mode, depending on the syringe capacity (0.1 ml/h increment)

flow rate accuracy ± 1% on mechanism ; ± 2 % on syringes. syringe capacities 5, 10, 20, 30/35, 50/60 CC

Types of syringe upto 50 types infusion modes : ml/h (0.1 ml/h increments). Volume/time: Volume to infuse 0.1 to 99.9 ml/Time: 0 h 01 to 96 h 00

• Volume limit: 0.1-999.9 ml (0.1 ml increments).priming 3 modes: mandatory, not mandatory, or advised/Rate: 1200 ml/h

• Bolus Manual bolus: Rate: 50 - 1200 ml/h (50 ml/h increment). end infusion (v/t & vl ) KVO 1 and KVO 2: from 0.1 to 5 ml/h adjustable and storable, continuous infusion or stop. fast start By default.

• Pause Programmable from 1 minute to 24 hours, increments from minute to minute.

• Night mode decreases the brightness of the screen and the green lights. The key beep can eventually be turned off. The night mode can be programmed manually or automatically in a variable time range.

• Drug libraries to be created with Vigilant, the IV Medication Safety Solution. • Adjustment of the dilution, authorised infusion modes, default infusion modes, flow rate and bolus limits in ml/h. upto 50 drugs.

Pressure management: 

• Pressure modes 2 modes available: variable or 3 pre-set levels - from 100 to 900 mmHg, 50 mmHg increment.

• DPS: The Dynamic Pressure System - warns of pressure variations

• A risk of obstruction or a possible leak in the infusion line can thus be anticipated

• Pressure monitoring Graphic representation of the pressure limit thanks to the pictogram

•Anti-bolus system Reduces significantly bolus after occlusion release (0.2 ml max.). 

Contact us

  • 50

    50 years of experience

  • Large distribution network

  • Quality technical services

More info

Alarms-pre-alarms-security:

• Pump status Green for infusion in progress, orange for pre-alarm and warnings, red for alarm. Visibility upto 5 meters.

•All alarms are expressed by means of light indicators, written words, pictograms and sound beeps

•Syringe position control Syringe barrel clasp check, plunger head detection, anti-siphon system check, flange detection 

• Infusion control Occlusion pressure pre-alarm, occlusion pressure alarm, end of infusion pre-alarm, end of infusion alarm

• Volume limit pre-alarm, volume limit alarm, keypad locking, hard and soft flow rate limits

Device control:

• Disengaged driving mechanism alarm, 

• Low battery pre-alarm, 

• Discharged battery alarm, 

• Battery capacity display in hours and minutes, 

• Unconfirmed programming, 

• Technical malfunction alarm (auto-test, rotation),

 Drive system advance check, 

• Watchdog check, 

• Communication connection failure

• Maintenance Preventive warning.

Technical specifications:

• Manual pusher Protection for the ongoing infusion thanks to “Push-Guard”

• Display Blue graphic LCD monochrome, size 66 mm x 33 mm (128 x 64 pixels)

• Swinglock clamp Versatile clamp that allows the fixation on a rail or on a pole (Pole: 20-40 mm max. / Rail: 25-35 x 10 mm)

• Stackability upto 3 devices self-stackable on a pole with a safe handle

• Dimensions (h/w/d) / weight 160 x 345 x 135 mm / 2150 g

•Battery LIFE NI/MH. Capacity minimum: 10 hours at 5 ml/h.

• Battery replacement: separate compartment – no need for electrical security testing

• Power suply 100 - 240 Vac / 15 VA, external power 9 Vdc by a converter.